14
Medical Gas Central Alarm - Medipoint 125 Digital Alarm
4233500107.03
Figure 13 - 2nd fix door installation.
2.2.3 Alarm panel front cover. Fit (see figure 13).
Secure the 2nd fix door to backbox flange by securing
hinge to backbox with screws provided. Plug in the
ribbon cable from the processor PCB to the power
supply PCB
No. Description
QTY
1
2nd Fix door (2nd fix supplied assembled)
1
2
M4x12 pan head (2nd fix, supplied loose)
2
3
M4 washer (2nd fix, supplied loose)
2
2.2.4 Earth Bonding.
Connection the earth cable assembly to the power
supply terminal block and earth stud on the backbox
and alarm door assembly. See figure 14.
Figure 14 - Earth bonding connections
No. Description
QTY
1
Earth cable (2nd fix, supplied assembled)
1
2
M3xNut (2nd fix, supplied loose)
4
3
M3 washer (2nd fix, supplied loose)
4
Typical earth stud assembly
2
2
3
1
Alarm Door
Earth Stud
Backbox
Earth Stud
Power Supply
Terminal Block
C
O
N
FI
D
EN
TI
AL
:
This
d
ocume
nt
is
ou
r p
ro
pe
rty
and
s
hal
l no
t w
ith
ou
t o
ur
pe
rm
issi
on
b
e
al
te
red
,
cop
ied
, u
se
d
fo
r m
anu
fa
ct
ur
ing
o
r c
om
m
un
ic
at
ed
to
an
y
ot
he
r pe
rs
on
o
r c
om
pa
ny
.
1
2
3
4
5
6
7
8
9
10
11
A
B
C
D
E
F
G
H
Parent 3D Model
8102371401_MP125_Open
Al
l materials
s
up
pli
ed
a
re
in
c
om
pl
ia
nc
e
w
ith
th
e
re
qu
ire
m
en
ts
o
f t
he
L
ist
o
f P
ro
hi
bit
ed
S
ubsta
nc
es
Title:
Material:
Finish:
Status
Work in Progress
REMOVE ALL BURR AND
SHARP EDGES UNLESS
OTHERWISE STATED
DO NOT SCALE
IF IN DOUBT ASK
---
As per order
DIMENSIONS IN MILLIMETERS
TOLERANCES UNLESS OTHERWISE
STATED
MACH. FAB.
0.25
1.0
0.1
0.05
0.5
0.5
1.0
-
NO DECIMAL PLACE:
ONE DECIMAL PLACE:
TWO DECIMAL PLACE:
ANGULAR:
SURFACE FINISH:
A3
DRAWING SCALE:
1 : 1.5
Third angle
DRAWING OWNER
SHEET:
PART NUMBER:
8102371401_MP125_
1
1 OF
PGC CODE:
GBNPBA
Revision history
Zone Rev
Description
IAR No.
Drawn By
Checked By
Approved
Date
1
2
3
Ribbon Cable connections