Medical Gas Central Alarm - Medipoint 125 Digital Alarm

4233500107.03

4. 

Commissioning

Before  starting  the  commissioning  actions  ensure 

that  all  points  from  Installation  section  2  and  alarm 

configuration section 3 have been completed.

Ensure all connected source equipment is operating in 

normal conditions with no alarm faults present.

The  object  of  commissioning  is  to  ensure  that  all 

components  are  serviceable  and  all  alarm  functions 

operate  satisfactorily.  The  commissioning  procedure 

also  ensures  that  anti-confusion  checks  are  carried 

out and that the correct gas service is displayed in the 

designated column.

 

WARNING!  Personnel  carrying  out  the 

following  commissioning  procedure  must  be 

qualified and fully conversant with the information 

contained in this manual.

 

WARNING! 

Before 

commencing 

the 

commissioning procedure ensure that all installation 

procedures  are  complete  and  that  all  wiring  is 

correctly  connected.  Before  switching  on  the 

mains electrical power supply, ensure the supply is 

correctly fused.

4.1 

Alarm panel test 

With  the  main  screen  displayed  check  that  all  gas 

services match the installation specification.  See figure 

35 for typical 6 gas unit with no faults.

Figure 35 - Example, 6 gas normal conditions.

Press the TEST switch and check all gas service indicator  

bars flash at the same rate, the audible warning sounds 

and the system fault LED flashes. 

Figure 36 shows examples of all status flashing.

See table 6 in section 5 “Operation of the Medipoint 125 

Digital alarm” for troubleshooting.

Figure 36 - Alarm Test Button, example 6 gas.

4.2 

Checking the slave relay output

If  the  slave  relay  is  to  be  used  in  the  installation, 

operation of the relay can be checked by ensuring that 

continuity exists between the relay terminals (N/O and 

C) when all NORMAL indications are illuminated. Check 

that  continuity  between  the  relay  contacts  is  broken, 

whenever  an  alarm  condition  occurs  during  tests  in 

section 

4.5 to 4.7

.

Note...

•  If the relay output is set to LCM, normal condition 

will be 180 Ohms and fault will be 510 Ohms.

•  If  the  relay  output  is  set  to  N/C  -  N/O,  normal 

condition  will  be  closed  and  fault  will  be  open 

contact.

4.3 

Checking  the  remote  audible 

warning devices.

Whilst  carrying  out  checks  described  in  section  4.5  to 

4.7, the opportunity should be taken to ensure that the 

remote audible warning operates whenever there is an 

alarm condition on it’s parent panel.

 

CAUTION!  The  TEST  and  MUTE  switches  must 

only  be  operated  by  gentle  finger  pressure.  Operation 

by the use of instruments, tools or other implements will 

cause damage to the switch and fascia.

4.4 

Medical gas alarm  faults.

The  procedure  to  check  the  alarm  fault  functions  is 

as  follows,  and  should  be  check  on  all  alarm  panels 

connected to the system:-

4.4.1  Alarm panel. Check NORMAL display. 

Ensure  all  connected  source  equipment  is  in  normal 

operation  with  no  faults  present.    The  alarm  should 

display  all  columns  as  green  normal,    power  on  LED 

illuminated  ,  system  fault  LED  off  and  no  audible 

Содержание 8102371401

Страница 1: ...Operation and Maintenance Instructions Medical Gas Central Alarm Medipoint 125 Digital Alarm Part number 4233500107 Revision 03 Jan 2021...

Страница 2: ...hone 44 0 1246 474242 Email gbn info beaconmedaes com Website Contacts www beaconmedaes com BeaconMed s reserves the right to make changes and improvements to update products sold previously without n...

Страница 3: ...in screen 3 6 Exit Setup Screen 3 7 Confi guration Menu Map 4 0 Commissioning 4 1 Alarm panel test 4 2 Checking the slave relay output 4 3 Checking the remote audible warning 4 4 Medical gas alarm fau...

Страница 4: ...ckaged and stored in controlled conditions Environmental Operating Conditions Adverse environmental conditions and harsh abrasives or chemicals may cause damage to the unit Environmental Protection Di...

Страница 5: ...rammed from on screen menus to provide a fully flexible alarm system Allowing the panels to be adapted to suit any hospital s medical gas monitoring requirements see section 1 12 for programming detai...

Страница 6: ...nt of any defect within the monitored functions the SYSTEM ALARM LED illuminates flashing accompanied by the audible warning and the defective gas service alarm display status bar illuminates An inter...

Страница 7: ...operty and shall not without our permission be altered ATLAS COPCO STANDARD Class Tolerances if not indicated according to Material Treatment Name Scale Family Drawing owner Secrecy Class 1102 K Compa...

Страница 8: ...p to four fault conditions yellow or red flashing alarm conditions to show the gas service status All alarm conditions are illuminated by a flashing display and accompanied by an audible warning Each...

Страница 9: ...on these PCB s which are interconnected by means of a multi way ribbon cable and polarised connector see figure 2 3 The area central alarm internal electrical installation complies with all relevant...

Страница 10: ...cular displayed alarm condition only This facility is designed for use when the plant or pipeline is shut off for a prolonged period On returning the pipeline pressure to NORMAL this feature automatic...

Страница 11: ...nnections have 5 terminals each See table 2 for details for standard options Each gas service and fault legend is also fully customisable through the on screen setup menus TABLE 2 GAS SERVICE CONNECTI...

Страница 12: ...uired by the contract specification Feed cables into the box leaving 400mm to 75 150 105 105 7 260 5 TOP 40 40 40 40 17 20 KNOCKOUT SEGMENTS 200 17 45 2 0 KNOCKOUT SEGMENTS BACKBOX SURFACE Figure 10 B...

Страница 13: ...Power supply PCB installation No Description QTY 1 Back Box 1st fix supply N A 2 Power supply PCB 2nd fix supplied loose 1 3 PCB Spacer 2nd fix supplied loose 4 4 M3 Nut 2nd fix supplied loose 4 5 M3...

Страница 14: ...nections No Description QTY 1 Earth cable 2nd fix supplied assembled 1 2 M3xNut 2nd fix supplied loose 4 3 M3 washer 2nd fix supplied loose 4 Typical earth stud assembly 2 2 3 1 Alarm Door Earth Stud...

Страница 15: ...igure 15 Alarm data connection MP125D 1 S1 ON IN OUT A B A B S MP125D 2 S1 OFF IN OUT A B A B S MP125D 3 S1 ON IN OUT A B A B S Note S terminal is for connecting the screen cable 2 2 6 Remote audible...

Страница 16: ...mmissionedimmediately disconnect the standby battery Connect electrical power supply wires to electrical power supply plug terminals E N and L 1 2 Figure 21 Alarm door security fastener No Description...

Страница 17: ...aracters Icons Icon Description Enter Key TAB key Moves to next field and highlights text Highlighted text will be overwritten on next key press Tick key Apply update and takes you to the main screen...

Страница 18: ...e Alarm setup 2 screen or press the Exit Icon to go back to the main operating screen 3 4 Alarm Setup Screen 1 Figure 26 The following steps describes the setup process for the alarm setup screen 1 3...

Страница 19: ...e new numerical password Press the TAB key or press the next field to go to the next field Confirm Use the num pad to enter the password again Press the ENTER key or press anywhere on the screen away...

Страница 20: ...setup process for the main gas service setup screen Figure 27 Main Gas Service Setup screen Special Characters Icons Icon Description Previous Icon Takes you to the previous menus screen Next Icon Tak...

Страница 21: ...k to set as inputs tick to set as remote transmission Press to toggle between the options 3 5 6 Location Special Characters Icons Description Exit Icon Takes you back to the main screen O2 Oxygen Gas...

Страница 22: ...elds will be editable 3 5 7 1 Standard gas options These fields allow for selecting the source equipment type available for the standard gas type selected Press on the matching equipment type to auto...

Страница 23: ...etup screen Figure 33 Main Gas Service Setup screen Special Characters Icons Icon Description Previous Icon Takes you to the previous menus screen Next Icon Takes you to the next menus screen Exit Ico...

Страница 24: ...key press Previous Icon Takes you to the previous menus screen Next Icon Takes you to the next menus screen Exit Icon Takes you back to the main screen Reject Icon Rejects changes Accept Icon Accepts...

Страница 25: ...6 shows examples of all status flashing See table 6 in section 5 Operation of the Medipoint 125 Digital alarm for troubleshooting Figure 36 Alarm Test Button example 6 gas 4 2 Checking the slave relay...

Страница 26: ...e table 2 section 1 8 for full list of standard gas service fault legends Check that the MUTE switch operates satisfactorily Press the audible alarm will be silenced and the mute icon will show on scr...

Страница 27: ...ED is flashing red The audible alarm is active See figure 39 fro example Figure 39 Example Gas service 1 connection fault 4 5 3 Input Reconnect Reconnect the input plug to gas channel System will retu...

Страница 28: ...d and the alarm condition remains the audible warning will re sound after 15 mins and a further MUTE selection will be required Should a further alarm condition occur after the panel has indicated a f...

Страница 29: ...mple shows gas service fault 4 Power LED Green On System Fault Black Off Gas Status Bar Red Flashing Audible Active Audible Mute Gas service fault flashing Mute activated Audible Silenced for 15 minut...

Страница 30: ...lank normal Yellow flashing fault Fault status 3 Blank Fault status 4 Blank normal Red flashing fault Manifold Control Systems Normal status Green normal blank during fault Fault status 1 Blank normal...

Страница 31: ...to the main screen automatically if a new alarm condition becomes active or after one minute if the alarm is not manually exited from the information screen 5 2 2 Data Logging Screen The data logging...

Страница 32: ...s distributors will be able to supply similar contracts in other areas CAUTION It is essential that only genuine BeaconMed ssparepartsarefittedduringmaintenance WARNING Obtain a work permit or equival...

Страница 33: ...33 Medical Gas Central Alarm Medipoint 125 Digital Alarm 4233500107 03 Figure 45a Alarm panel assembly details...

Страница 34: ...34 Medical Gas Central Alarm Medipoint 125 Digital Alarm 4233500107 03 Figure 45b Alarm panel assembly details...

Страница 35: ...e 5 year maintenance replace the LCD screen and backup butteries See maintenance procedures in 6 1 and figures 45a 45b for details 8 RECOMMENDED SPARES Table 8 details the list of recommended spares t...

Страница 36: ...2460 Atlas Copco Medical Ltd Telford Crescent Staveley Derbyshire S43 3PF UK Tel 44 0 1246 474242 www beaconmedaes com Atlas Copco Airpower n v Boomsesteenweg 957 2610 Wilrijk Belgium...

Отзывы: