Medical Gas Central Alarm - Medipoint 125 Digital Alarm
4233500107.03
This will take you to the Password Screen. See figure 25.
Figure 25 - Password Screen
Use the num-pad to enter the 4 digit numerical password
(Default: 1234).
Press the ENTER
key to confirm the password and
take you to the setup menus.
Settup menus start on the Alarm setup 1 screen.
Figure 26 - Alarm setup 1 screen
Special Characters/Icons:
Icon Description
Previous Icon.
Takes you to the previous menus screen.
Next Icon.
Takes you to the next menus screen.
Exit Icon.
Takes you back to the main screen.
Reject Icon.
Rejects changes.
Accept Icon.
Accepts Changes
From the “Alarm setup 1” screen click the “Next Icon”
to go to the “Gas service setup” screen, press the
“Previous Icon”
to go to the “Alarm setup 2” screen, or
press the “Exit Icon”
to go back to the main operating
screen.
3.4
Alarm Setup Screen 1 (Figure
26).
The following steps describes the setup process for the
alarm setup screen 1.
3.4.1 Date:
Press on the first field (Day), the num-pad will appear.
Hold down to highlight current text ready to overwrite.
Use the num-pad to enter the day.
Press the TAB
key to go to the next field (Month). Use
the num-pad to enter the month.
Press the TAB
key to go to the next field (Year). Use
the num-pad to enter the year.
Press the ENTER
key or press anywhere on the screen
away from the NUM-PAD to remove the NUM-PAD.
3.4.2 Time.
Press on the first field (Hours), the num-pad will appear.
Hold down to highlight current text ready to overwrite.
Use the num-pad to enter the hours.
Press the TAB
key to go to the next field (Minutes).
Use the num-pad to enter the minutes.
Press the TAB
key to go o the next field (Seconds).