BDDF00001 Issue 5
53/56
Alaris™ Syringe Pumps
Trumpet Curves and Start-up Curves
Alaris GH Syringe Pump and Alaris GH Guardrails Syringe Pump
Start-up Trend . BD Plastipak 5ml @ 0 .1ml/h
Trumpet Curve . BD Plastipak 5ml @ 0 .1ml/h
Ra
te (ml/h)
0.0
0.1
0.2
0
20
40
60
80
100
120
Er
ror (%)
-30.0
-20.0
-10.0
0.0
10.0
20.0
30.0
0
5
10
15
20
25
30
35
Time (mins)
Observation Window (mins)
Maximum Error
Minimum Error
Linear Mean = -2 .5%
Start-up Trend . BD Plastipak 50 ml @ 1 .0 ml/h
Trumpet Curve . BD Plastipak 50 ml @ 1 .0 ml/h
Ra
te (ml/h)
Er
ror (%)
Time (mins)
Observation Window (mins)
Maximum Error
Minimum Error
Linear Mean = -1 .8%
Start-up Trend . BD Plastipak 50 ml @ 5 .0 ml/h
Trumpet Curve . BD Plastipak 50 ml @ 5 .0 ml/h
Ra
te (ml/h)
Er
ror (%)
Time (mins)
Observation Window (mins)
Maximum Error
Minimum Error
Linear Mean = +0 .2%