Alaris™ Syringe Pumps
Alarms and Warnings
Display
Alarm
Priority
Description and Troubleshooting Guide
ADD DRUG NOT COMPLETE
Low
The
ADD DRUG
operation has not been completed after five seconds the User will be
notified by an auditory cue . After two minutes have expired and the
ADD DRUG
operation
has not been completed, a low priority alarm is generated . Press
CANCEL
softkey and
complete
ADD DRUG
operation .
ATTENTION
Low
If the Pump has been left on for more than 2 minutes* (referred to as
ATTENTION
in the
log) without starting the operation, a low priority alarm will be generated . Press the
R
button to silence the alarm for a further 2 minutes . For extended Attention Timeout press
and hold down the
R
button and wait for four beeps in succession, this will put the
Pump on standby for 15 minutes .
*Configurable option .
Note:
The audio sound pressure level is at least 45 dB depending on configuration of the alarm sound level .
w
Setting the alarm sound pressure level lower than the ambient sound pressure level can impede User recognition of
alarm conditions.
Alarm Priority Level Indicators
Priority
Audio Indicator
Visual Indicator (Beacon)
HIGH
Ten beep sequence followed by a three second pause
Flashing Red
MEDIUM
Three consecutive beeps followed by a four second pause
Flashing Amber
LOW
Three consecutive beeps followed by a sixteen second pause
Solid Amber