BDDF00001 Issue 5
31/56
Alaris™ Syringe Pumps
Prompts
Prompts
Prompts are indicated by an audible alarm and message, they cannot be silenced and do not have a visual indicator .
Display
Icon
Description and Troubleshooting Guide
DOSE WOULD EXCEED*
The dose rate has been set to a value which exceeds a Guardrails Soft Alert . Check
infusion setting, to continue with infusion at set rate confirm
OVERRIDE LIMIT
by
pressing the
YES
softkey . If
OVERRIDE LIMIT
is not required press the
NO
softkey and
adjust rate below Guardrails Soft Alert .
DOSE UNDER*
The dose rate has been set to a value which is under a Guardrails Soft Alert . Check
infusion setting, to continue with infusion at set rate confirm
OVERRIDE LIMIT
by
pressing the
YES
softkey . If
OVERRIDE LIMIT
is not required press the
NO
softkey and
adjust rate above Guardrails Soft Alert .
DOSE NOT PERMITTED
The dose rate has been set above a Hard Limit . Check infusion setting and adjust rate
to appropriate required rate .
BOLUS DOSE OVER*
The bolus dose has been set to a value which exceeds a Guardrails Soft Alert . Check
the bolus setting, to continue with the bolus confirm
OVERRIDE LIMIT
by pressing the
YES
softkey . If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust dose
below Guardrails Soft Alert .
BOLUS DOSE UNDER*
The bolus dose has been set to a value which is under a Guardrails Soft Alert . Check
the bolus setting, to continue with the bolus confirm
OVERRIDE LIMIT
by pressing the
YES
softkey . If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust dose
above Guardrails Soft Alert .
BOLUS DOSE NOT PERMITTED
The bolus dose has been set above a Hard Limit . Check bolus setting and adjust to
appropriate required dose .
CONCENTRATION NOT PERMITTED
The drug concentration has been set above or below a Hard Limit . Check the amount
and total volume and adjust to give the appropriate required concentration .
WEIGHT OUTSIDE LIMIT*
The patient weight has been set to a value which exceeds or is under a Guardrails Soft
Alert . Check the weight setting, to continue confirm
OVERRIDE LIMIT
by pressing the
YES
softkey . If
OVERRIDE LIMIT
is not required press the
NO
softkey and adjust the
value within the limits .
RATE NOT PERMITTED
The infusion rate has been set above a Hard Limit . Check infusion setting and adjust to
appropriate required rate .
*
Guardrails