BDDF00001 Issue 5
26/56
Alaris™ Syringe Pumps
Pressure Features
Pressure Features
Auto Set Pressure (If enabled)*
If the Auto Set Pressure Option is enabled then the Pump
automatically
adjusts the pressure occlusion limit .
After 15 minutes of infusion the Pump
automatically
adjusts the occlusion pressure limit to XX mmHg above the average infusion
pressure, taken from the average of the last five minutes of infusion .
Note:
XX is the
AUTO OFFSET
pressure and is determined by the user . This adjustment,
AUTO OFFSET
value 15-100mmHg, is
configurable by profile within the data set . At pressures up to 100mmHg the
AUTO OFFSET
value is added . For pressures
above 100mmHg the alarm level is set to whatever the
AUTO OFFSET
value is as a percentage above the average infusion
pressure up to the maximum pressure defined within the data set .
*
Pressure Level with pressure set fitted*
1 . To check and adjust the pressure level press the
e
button . The display will change to show a 20 minute pressure trend graph
displaying the pressure alarm level and the current pressure level .
2 . Press the
f
keys to increase or decrease the pressure alarm level . The new level will be indicated on the display .
3 . The
AUTO
Pressure feature may be used when a stable pressure has been achieved over a short period of infusion . If
AUTO
Pressure
has been enabled the automatic pressure alarm level is calculated and set by pressing the
AUTO
softkey .
4 . Press the
TREND
softkey to view the pressure trend of the previous 12 hours . The pressure trend can be viewed at 15 minute
intervals by using the
+/-
softkeys . The pressure trend graph displays the pressure at a given time .
5 . Press the
OK
softkey to exit the pressure screen .
_
+
OK
20mmHg
07:45
*
Pressure Level*
1 . To check and adjust the pressure level press the
e
button . A bar graph will be displayed showing the pressure alarm level and the
current pressure level .
2 . Press the
f
keys to increase or decrease the alarm level . The new level will be indicated on the display .
3 . Press
OK
to exit the screen .
w
The interpretation of the pressure readings and occlusion alarms are the responsibility of the clinician and should
include the clinical context in which the Pump is being used.
*
without pressure set fitted (not applicable when Fully Dedicated)