BDDF00001 Issue 5
39/56
Alaris™ Syringe Pumps
Specifications
Dimensions
335 mm (w) x 121 mm (h) x 200 mm (d) .
GH
310 mm (w) x 121 mm (h) x 200 mm (d) .
Weight
2 .4 kg (excluding power cable) .
Protection against fluid ingress
IP32 - Protected against direct sprays of water up to 15° from vertical and protected against solid objects greater than 2 .5mm .
Note:
IP33 applies if mains retainer kit, part number 1000SP01294, is fitted .
Alarm Conditions
Drive Disengaged
Occlusion
Attention (Nurse Callback)
Check Syringe
Battery Low
Titration not confirmed
Line Occlusion
Battery Empty
VTBI Done
Near End Of Infusion
End of Infusion
AC Power Fail
Internal Malfunction
Pressure Disc Out
Dose Under
Dose Would Exceed
Dose not Permitted
Bolus Dose not Permitted
Bolus Dose Under
Bolus Dose Over
Rate not Permitted
Concentration not Permitted
Weight Outside Limit
Add Drug Not Complete
Environmental Specifications
Operating Temperature
0°C - +40°C
Operating Relative Humidity
20% - 90%
Operating Atmospheric Pressure
700hPa - 1060hPa
Transport & Storage Temperature
-30°C - +50°C
Transport & Storage Relative Humidity
10% - 95%
Transport & Storage Atmospheric Pressure
500hPa - 1060hPa
Electrical/Mechanical Safety
Complies with EN/IEC60601-1 and EN/IEC60601-2-24 .
Potential Equalisation Conductor
The function of the Potential Equalisation Connector (Conductor) is to provide a direct connection between the Pump and the potential
equalisation busbar of the electrical installation . To use the Potential Equalisation Connector, connect the Potential Equalisation
Connector on the Pump to the potential equalisation busbar of the electrical installation .
EMC
Complies with EN/IEC60601-1-2 and EN/IEC60601-2-24 .