BDDF00001 Issue 5
36/56
Alaris™ Syringe Pumps
Configured Options
Alaris Editor Software Profile Drug library
The following drug parameters are only configurable via the Alaris Editor Software, see
Alaris Editor Directions For Use
for details on how
to configure Profile Drug Library, and are used when the Pump is operated with a drug name selected .
Concentration Units
The unit for concentration parameters
Concentration Limits
(Min and Max)
These define the range over which the drug concentration can be modified
during programming of the Pump .
Continuous Dose Rate -
Units
The continuous dose rate units . Can be based on patient weight .
Soft Min*
The continuous dose rate value below which override confirmation is required .
Default
The default continuous dose rate offered when the drug is selected .
Soft Max*
The continuous dose rate value above which override confirmation is required .
Hard Max
The maximum allowed continuous dose rate .
Bolus Mode
Bolus feature can be set to one of the following options:
• Disabled
• Hands-On only
• Hands-Free and Hands-On
Bolus Dose -
Units
The bolus dose units . Can be based on patient weight .
Soft Min*
(Hands-Free only)
The bolus dose value below which override confirmation is required .
Default
(Hands-Free only)
The default bolus dose offered .
Soft Max*
(Hands-Free only)
The bolus dose value above which override confirmation is required .
Hard Max
The maximum allowed bolus dose .
Bolus Rate -
Default
The default value for bolus rate in ml/h .
Occlusion Alarm Pressure
GH
The default occlusion alarm pressure .
Occlusion Alarm Pressure
The default occlusion alarm pressure .
*
Guardrails
Profile Syringe Library
The Profile Syringe Library is created from the predefined Master Syringe Library .
Check the boxes of the syringes to be included in the profile . Checking the
All Syringes
checkbox under
Operations
selects all the
syringes .
For Syringe brands and sizes available see 'Recognised Syringes' section .
Note:
It is recommended that only syringe types and sizes used in the care area are selected .