8
|
Functions
Baumer
34
Operating Manual
OM20/ OM30 RS485 with Modbus RTU | V1
Invalid Value Handling – example 2
n
Parameters:
▪ Time period (Hold Time):
1000 ms
n
Interpretation: After expiration of the time period, the placeholder for an invalid value is out-
put at the digital output.
Digital output
50
250
1
2
3
4
5
s
mm
Hold Time
Signal quality
NaN
Ill. 10:
Invalid Value Handling – example 2
Modbus RTU command – holding register: Invalid Value Handling
Address 800 – Dropout Timeout:
Address
Access
Number of
registers
Data type
Description
800
Read/write
1
uint16_t
Interval for suppressing invalid mea-
sured values [ms]