SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
40
Write to Disk
-Check the
“
Write Selected Data At Every
1
Frequency
Settings
”
, and configure the file name and folder to store the specific
wavelength intensity by fixed time interval. You need to complete
“
Configure
Time Acq. Channel
” also. Please refer the step
⑦
to
configure the channels.
-Check the “
Write All Spectrum At every
1
Frequency Settings
”
,
and configure the file name and folder to store the full spectrum by
fixed time interval.
If you only want to store the full spectrum, there is no need to set up
the “
Configure
Time Acq. Channel
”.
Initial Delay
Set the start time of the measurement (After pushing the play button).
The measurement will start after
“
Initial Delay
”
time.
Frequency
Set the time interval. Larger than 100 ms.
Duration
Set the period of measurement.
③
Write the folder and file name for the time series acquisition process.
④
Enter the “
Initial Delay
” to set starting time.
⑤
Enter the “
Frequency
” to set the data collection frequency.
⑥
Enter the “
Duration
” to set the data collection period.
⑦
Open “
Menu
”/”
Time
”/”
Configure
Time Acquisition Channel
”, or click the
icon. In this window you can configure the wavelength channels. Six specific
wavelengths can be set up from channel 1 to channel 6.
Содержание SEC2000 Spectra System
Страница 1: ......
Страница 61: ...SEC2000OperatingManual SEC2000 Spectra System www als japan com 61 Fig 7 2 5...
Страница 66: ...SEC2000OperatingManual SEC2000 Spectra System www als japan com 66...