SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
56
②
A reference spectrum and a dark spectrum are measured in the same procedure
as absorbance measurement. Put a standard sample into the cuvette holder in
absorbance mode, the screen will become as Fig. 6-2-9. The following
procedures make 1 mM potassium ferricyanide as an example.
<Fig. 6-2-9> Absorbance of 1 mM K
3
Fe(CN)
6
solution.
③
Open the “
Quantitative Analysis
” window by “
View
”/“
Spectrum Tool Bar
”/
check the “
Quantitative Analysis
”. The window of “Quantitative Analysis” is
shown on the right side of the main screen (Fig. 6-2-9).
④
Set parameters (Fig. 6-2-10)
Set the file name and folder of saving calibration curve by “
Save
”.
Set the peak wavelength of the sample by “
Wavelength_ nm
”, we set 413 nm
for the ferricyanide.
Set the unit of solution concentration by “
Con.
”, we choose “
mM
”.
Set the unit of absorbance by “
Intensity
”, we choose “
O.D.
”.