SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
37
lamps are stabilized.
③
Open the software and click “
Go
” icon, the spectrum signal will appear.
<Fig. 5-6-1> Configuration example of relative irradiance measurement.
④
“
Spectrum
”/”
Reference Color Temp.
” Input the color temperature (K) of the
standard light source used for relative irradiance measurements.
⑤
On Scope Mode, Adjust the Integration Time to control the maximum value of the
Y axis of the graph (less than 16,000).
⑥
By clicking the “
Reference
” icon in the program, the spectrum is obtained at the
maximum light intensity. Save the reference spectrum by “
File
”/ ”
Save
”/
“
Reference
”.
⑦
Close the light source shutter (Do not close the power switch), and click the “
Dark
”
icon in the program to obtain a spectrum under the minimum conditions without
light. Save the dark spectrum by “
File
”/”
Save
”/”
Dark
”.
⑧
Connect a luminescence sample to integrating sphere (set a LED bulb to
integrating sphere), Click the “
Irradiance Mode
” icon to change the Mode.
<Fig. 5-6-2> Relative irradiance
measurement of LED bulb.