SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
29
4-3-2. Functions
①
Icons
Minus
Dark
Select this icon if the baseline of the resulting graph is unstable
based on the present Y coordinate, or to adjust the Y coordinate
based on the dark level. If this icon is clicked, the baseline will
be forced to be Y (Intensity) = 0 which will be reflected in the
graph. Though this function is used for adjusting an unstable
distribution of a spectrum, you must pay attention to the fact that
it is based on the original data when saved as a file.
Auto
Scale
For a detailed view, select this button to enlarge the Y-
coordinate into full screen. The X-coordinates may be enlarged
beyond the edge of the screen in this mode.
Full
Scale
This icon displays the X-axis and Y-axis in a full screen.
Scale
For a detailed view, this mode allows the user to examine the graph
with the X-axis and Y-axis scaled to the user’s preference. Simply
input the upper range of each axis accordingly.
Enable
This function provides the coordinate of a particular peak or
point. After choosing
“ENABLE”
icon, select the target point.
Then the result will be recorded on the point that the mouse
indicates.
Cursor
With this function selected, clicking on a point will provide the
coordinate information (Wavelength, Intensity) shown on the
status bar at the bottom of the screen. In accordance with the
movement of the mouse, new coordinate information will be
shown.