SEC2000 Operating Manual
SEC2000 Spectra System
www.als-japan.com
54
⑥
On the step of ”
Quantitative Analysis- Calibration
”, to measure the absorbance of
the standard material with known concentration, put a standard sample into the
cuvette holder, and input data in “
Concent.
” Column. Move your mouse to
“
Absorb.
” column, and click “
Get Absorb.
” icon to calculate the area of the
standard sample (get a calibration line with at least three different concentrations of
standard material).
<Fig. 6-2-6> Absorbance of K
3
Fe(CN)
6
solutions (Concentration: 1, 0.5 and 0.25 mM).
⑦
Click “
Use
”, then calibration line is obtained as shown in Fig. 6-2-7.
<Fig. 6-2-7> Calibration line with three different concentrations.
The standard samples are shown in closed circle.