9
Locating your EASYpure RO
Your EASYpure RO must be located within six feet
of an electrical outlet appropriate for your unit, with-
in five feet of your feedwater supply and close to
an atmospherically vented drain. If you are using
the accessory 30 Liter Reservoir (Part Number
TY742X2A), the tank overflow on the EASYpure
RO's integral 6.5 liter tank should be located below
the level of the storage reservoir's overflow. This
will ensure that the EASYpure RO's internal float
switches will control the operation of the system. If
the EASYpure RO is to be used with the accesso-
ry 30 Liter Reservoir, the reservoir must not be
mounted farther than 4 feet away from the
EASYpure RO. The EASYpure RO Bench Stand
(Part Number D7424) will be required for a bench
mounted installation with an accessory 30 Liter
Reservoir.
General usage
Do not use this product for anything other than its
intended usage.
Do not mount the EASYpure
RO directly over equipment
that requires electrical ser-
vice. Routine maintenance of
this unit may involve water
spillage and subsequent
electrical shock hazard if
improperly located.
Warning
Installation