Wall Mounting
If you are wall mounting the EASYpure RO using
the optional wall mounting bracket, mount the unit
as follows:
Wall Mounting the EASYpure RO
1.
Install the wall bracket (Part Number
D7384) on the wall in a location that is
accessible to water, power and an atmos-
pherically vented drain, and that is conve-
nient to use. A minimum of 4 fasteners
must be used.
2.
Remove the 4 feet from the EASYpure
RO. Retain the screws.
3.
Place the EASYpure RO on the wall
bracket swivel base so that the screw
holes where the feet were attached line up
with the holes in the wall bracket. There
are guides on the wall bracket that will
mate with the EASYpure RO.
I
NSTALLATION
Figure 2 Wall Mounting the EASYpure RO
Wall composition and con-
struction, as well as fastener
type, must be considered
when mounting this unit. The
mounting surface and fasten-
ers selected must be capable
of supporting a minimum of
120 lbs.; inadequate support
and/or fasteners may result in
damage to mounting surface
and/or equipment. If you are
unsure of mounting surface
composition, condition and
construction, or correct fas-
teners, consult your building
maintenance group or con-
tractor.
Do not mount the EASYpure
RO directly over equipment
that requires electrical ser-
vice. Routine maintenance of
this unit may involve water
spillage and subsequent elec-
trical shock hazard if improp-
erly located.
Note
If you are bench mounting the
unit, proceed to Utility
Connections.
Warning
Содержание EASYpure D7421
Страница 2: ...2 Figure 1 EASYpure RO Ventgard Back Door Latch Pressure Gauge Indicator LED s Tank Water Level Sight...
Страница 39: ...39 Wiring Diagrams Figure 14 Wiring Diagram 100 Volt...
Страница 40: ...40 WIRING DIAGRAMS Figure 15 Wiring Diagram 120 Volt...
Страница 41: ...41 WIRING DIAGRAMS Figure 16 Wiring Diagram 230 Volts...
Страница 42: ...42 EASYpure RO Exploded View...