18
2.
Remove an RO membrane car-
tridge (Part Number D7425) from
its plastic bag.
3.
Remove the nut, stainless steel
grab ring, backup ring and o-ring
from the connectors on the mem-
brane.
4.
Connect the product and reject
tubes in the right-hand membrane
position on the EASYpure RO to
the product and reject outlet con-
nectors on the RO membrane car-
tridge. Ensure that the o-ring, back-
up ring, stainless steel grab ring
and nut are in place on the product
and reject tubing on the EASYpure
RO (see Figure 4, page 12).
Ensure that the product tube is
connected to the product connector
and the reject tube is connected to
the reject connector. Hand-tighten
the connections.
I
NSTALLATION
Figure 6 Proper Position of Prefilter and Membranes
Prefilter
Right-hand
Membrane
Position
Left-hand
Membrane
Spring Metal
Membrane Retainer
Do not use a wrench on the
tubing adapter connections.
A sufficiently tight connection
can be made by hand.
Caution