33
Fuse Replacement
1.
Disconnect the EASYpure RO from its
power source. Remove the outer cover.
2.
Locate the fuse holder, part of the power
entry module (see Figure 13 for location),
and remove.
3.
Remove the burned out fuse(s) and
replace with one(s) of equal size and rat-
ing.
Membrane Replacement
A variety of factors will determine the service life of
your EASYpure RO membranes. With proper main-
tenance, the membranes should last approximately
one year for normal usage.
When either the "Service Membrane" or the
"Replace Membrane" Indicator lights, use the fol-
lowing procedure to replace the membranes:
1.
Disconnect the EASYpure RO from its
power source.
2.
Remove the membranes and discard.
S
ERVICE AND
R
EPAIR
Fuse Holder
Location
Figure 13 Fuse Holder Location
Warning
Replace fuses with those of
the same type and rating.