13
Reject Water Connection
1.
Locate the reject water tubing. This is the
piece of 3/8" O.D. tubing that has an
adapter on one end and no connector on
the other end. Remove the nut, grab ring,
backup ring and o-ring from the reject
water fitting located at the lower right of
the unit. (See Figure 4 for part identifica-
tion and Figure 3 for fitting location.)
2.
Thread the nut, grab ring and backup ring
on the reject adapter. Do not use the o-
ring at this time.
3.
Push the tubing through the hex nut until it
bottoms out in the adapter.
4.
Remove the adapter nut and tubing. Place
the o-ring over the tubing. Be careful not
to push the backup ring or grab ring fur-
ther back on the tubing when installing the
o-ring. Connect the tubing to the reject
connection on the EASYpure RO.
I
NSTALLATION
Figure 4 Typical Polypropylene Tubing Adapter Installation