Calibrating Barco CRT displays
1 If you use an X-Rite DTP92 sensor, you may be asked to calibrate the
sensor first.
Make sure the sensor is completely covered (so that no light can enter
the sensor) and click on OK.
2 Put the sensor, with Ambient Light Shield if appropriate, on the image
on the screen.
Important:
Make sure there are no application windows or other items (e.g., the
cursor) displayed under the sensor. They would cause the calibration
to fail.
When ready, click on OK.
3 For sensors that have an ambient light shielding only: If the Ambient
Light Compensation option in the Preset is checked, the software asks
you to remove the sensor from the screen and put it back without
Ambient Light Shield.
When ready, click on OK.
After a few measurements, the software asks to remove the sensor
once more, and put it back with the Ambient Light Shield.