101
Clicking on the Contrast (Brightness) icon left of the sliders, puts the
Contrast (Brightness) control in calibrated position. This is the preferred
position.
Luminance Uniformity Correction on/off
This function is available for BARCO displays only.
In the Configuration section in the main window, right-click on the icon of
the display of which you wish to switch LUC on or off.
From the drop-down menu that appears, select Luminance Uniformity
Correction...
The Luminance Uniformity Correction system, if calibrated by means of
“Full Screen Calibration” (see below), ensures a uniform luminance, even
in the corners of the picture tube.
This function is grayed if the function is not available on the display.
Full Screen Calibration
This function starts with a normal calibration in the center of the
screen. It proceeds with calibration in a number of zones over the picture
tube.
Please refer to § Full screen calibration : “ Full screen calibration “ for
more information.
The function is grayed if not available on the display.
Degaussing displays
This function is available for color BARCO displays only.
In the Configuration section in the main window, right-click on the icon of
the display you wish to degauss.
Содержание MediCal Pro
Страница 1: ...Installation User Manual MediCal Pro...
Страница 2: ...www barco com B4100132 02 July 2004...
Страница 9: ...Introduction 9 INTRODUCTION...
Страница 18: ...Introduction 18...
Страница 19: ...First installation 19 FIRST INSTALLATION...
Страница 31: ...First installation 31 7 The entered information is added to MediCal Administrator Click on Next...
Страница 65: ...65 USING MEDICAL PRO...
Страница 104: ...104 About Displays the details of this version of MediCal Pro...
Страница 105: ...105...
Страница 106: ...106 GLOSSARY...
Страница 110: ...110...