Cleaning Instructions
6
Nio 2-3-5 MP
Cleaning the glass plate- Outside
surface
Proceed as follows:
1. Take one of the nonabrasive cloths and put glass cleaner liquid on
it.
2. Clean the glass with the moist cloth.
3. Take another clean, dry cloth and wipe the glass dry.
Cleaning the glass plate- Inside
surface
If there are dust particles on the inside of the glass plate, proceed as
follows:
Spray with the duster gas onto the surface to be cleaned. For best
results, use the 'quick shot' method (Spray in short blasts). When the
aerosol is cooling down and the pressure is decreasing (internal
evaporation to restore the equilibrium in the aerosol can), wait a few
moments for the internal pressure to restore and repeat the spraying
action. Do not shake or move can during use.
Spraying the gas, the dust particles are blown away from the glass.
Do not blow them in the direction of the display.
Attention
: The dust remover can contains a liquid gas. If you shake the
can or move the can too fast while spraying, you may blow drops of
liquid on the glass surface!
If this is the case, clean the glass as described in the § "Outside
surface" above.
Cleaning the LCD panel
Proceed as follows:
1. If the LCD panel is dirty, take one of the nonabrasive cloths and
moisten the cloth with a solution of 25% Isopropyl Alcohol (IPA)
and 75% de-ionized or distilled water.
Do not use the same cloth as the one you apply glass cleaner to
for cleaning the glass.
2. Clean the panel with the moist cloth.
3. Take another clean, dry cloth and wipe the glass dry.
4. If there are dust particles on the LCD panel, blow them away by
using the dust remover. Do not blow them in the direction of the
glass plate.
Attention
: The dust remover contains a liquid gas. If you shake the
can or move the can too fast while spraying, you may blow drops of
liquid on the panel surface!
If this is the case, clean the panel as described above.
Cleaning the plastics
Clean the cabinet using a recognized cleaning product for medical
equipment. The cloth you use must be moist, not wet!
The cabinet has been tested for resistance to the following products:
Cidex, Betadine, Alcohol (Isopropyl and Ethyl), Ammonia-based
cleaners (Windex) and Aquasonic Gel.
Содержание E-2320 C
Страница 2: ...2 Nio 2 3 5 MP...
Страница 11: ...First level service Nio 2 3 5 MP 11 5 First level service...
Страница 18: ...Advanced service 18 Nio 2 3 5 MP 6 Advanced service...
Страница 34: ...BLOS calibration 34 Nio 2 3 5 MP Check the Include Backlight check box Click the Calibrate button...
Страница 35: ...Functional tests Nio 2 3 5 MP 35 7 Functional tests...
Страница 43: ...Packaging instructions Nio 2 3 5 MP 43 18 Close the box and tape it...
Страница 49: ...Technical specifications Nio 2 3 5 MP 49...