BLOS calibration
Nio 2-3-5 MP
33
Remarks:
•
If the ambient temperature changes, redo the zero calibration
The zero calibration can be checked in normal working position by
putting the measurement head on a flat surface, so no light will reach
the sensor and the display will show 0.00 cd/m². If the zero calibration
is not well performed, the message “offset error” will appear or a
value greater then zero will be measured.
Calibrate the BLOS
•
Connect the metal calibration cover to the display.
•
Connect the CA 110 / 210 to the PC by a proper cable (see illustra-
tion) and put it in the calibration cover.
•
Start the program
Mfdcontrol.exe
(version Sept. 29 2003 at least).
•
In the main menu, click the
I-Guard
button
•
Click on
Monochrome I-Guard
•
In the next menu, the program will detect the BLOS.