K5902079-07
Coronis Uniti
41
To select a display function:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration
menu.
3.
Enter the
Display Function
submenu.
4.
Select one of the available display functions and confirm.
5.19 Ambient Light Compensation (ALC)
About ALC
Ambient Light Compensation (ALC) can only be enabled on your display when the DICOM display
function is selected. Therefore, please refer to
“
”
, page 40 to correctly set the
display function.
When ALC is enabled, the DICOM display function will be recalculated taking a preset ambient light correction
value into account. This value is determined by the selected reading room. Therefore, it is also important to
select a realistic reading room when enabling ALC. This can be done by following the instructions in
, page 41.
To enable/disable ALC:
1.
Bring up the OSD main menu.
2.
Navigate to the
Configuration > Calibration > Ambient Light
menu.
3.
Enter the
Ambient Light Compensation
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
5.20 Reading rooms
About reading rooms
Reading rooms can only be selected when the DICOM display function is selected. Therefore,
please refer to
“
”
, page 40 to correctly set the display function.
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
Ambient Light Compensation (ALC)
, page 41 to enable ALC.
The available reading rooms for your Coronis Uniti are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in office rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
Содержание Coronic Uniti
Страница 1: ...ENABLING BRIGHT OUTCOMES MDMC 12133 User Guide Coronis Uniti...
Страница 5: ...5 K5902079 07 Coronis Uniti Welcome 1...
Страница 7: ...7 K5902079 07 Coronis Uniti Parts controls and connectors 2...
Страница 12: ...K5902079 07 Coronis Uniti 12 Parts controls and connectors...
Страница 27: ...27 K5902079 07 Coronis Uniti Daily operation 4...
Страница 33: ...33 K5902079 07 Coronis Uniti Advanced operation 5...
Страница 48: ...K5902079 07 Coronis Uniti 48 Advanced operation...
Страница 49: ...49 K5902079 07 Coronis Uniti Cleaning your display 6...
Страница 51: ...51 K5902079 07 Coronis Uniti Repackaging instructions 7...
Страница 58: ...K5902079 07 Coronis Uniti 58 Repackaging instructions...
Страница 59: ...59 K5902079 07 Coronis Uniti Important information 8...
Страница 73: ...K5902079 07 Coronis Uniti 73 Important information...