K5902079-07
Coronis Uniti
65
This Coronis Uniti complies with appropriate medical EMC standards on emissions to, and interference from
surrounding equipment. Operation is subject to the following two conditions: (1) this device may not cause
harmful interference, and (2) this device must accept any interference received, including interference that
may cause undesired operation.
Interference can be determined by turning the equipment off and on.
If this equipment does cause harmful interference to, or suffer from harmful interference of, surrounding
equipment, the user is encouraged to try to correct the interference by one or more of the following measures:
•
Reorient or relocate the receiving antenna or equipment.
•
Increase the separation between the equipment and receiver.
•
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
•
Consult the dealer or an experienced technician for help.
Electromagnetic immunity
The Coronis Uniti is intended for use in the electromagnetic environment specified below. The customer or the
user of the Coronis Uniti should assure that it is used in such an environment.
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment
–
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15
kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the
relative humidity should be
at least 30%
Electrical fast transient/
burst
IEC 61000-4-4
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
± 2 kV for power supply
lines
± 1 kV for input/ output
lines
100 kHz repetition
frequency
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Line to line: ± 0.5 kV, ± 1
kV
Line to ground: ± 0.5 kV, ±
1 kV, ± 2 kV
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
0% residual voltage for 0.5
period at 0°, 45°, 90°,
135°, 180°, 225°, 270°
and 315°
0% residual voltage for 1
period at 0°
70% residual voltage for
25 periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user of
the Coronis Uniti requires
continued operation
during power mains
interruptions, it is
recommended that the
Coronis Uniti be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m
Not applicable
5
Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
5:
Coronis Uniti doesn
’
t contain susceptible components to magnetic fields
Содержание Coronic Uniti
Страница 1: ...ENABLING BRIGHT OUTCOMES MDMC 12133 User Guide Coronis Uniti...
Страница 5: ...5 K5902079 07 Coronis Uniti Welcome 1...
Страница 7: ...7 K5902079 07 Coronis Uniti Parts controls and connectors 2...
Страница 12: ...K5902079 07 Coronis Uniti 12 Parts controls and connectors...
Страница 27: ...27 K5902079 07 Coronis Uniti Daily operation 4...
Страница 33: ...33 K5902079 07 Coronis Uniti Advanced operation 5...
Страница 48: ...K5902079 07 Coronis Uniti 48 Advanced operation...
Страница 49: ...49 K5902079 07 Coronis Uniti Cleaning your display 6...
Страница 51: ...51 K5902079 07 Coronis Uniti Repackaging instructions 7...
Страница 58: ...K5902079 07 Coronis Uniti 58 Repackaging instructions...
Страница 59: ...59 K5902079 07 Coronis Uniti Important information 8...
Страница 73: ...K5902079 07 Coronis Uniti 73 Important information...