Appendix A. Outline and Installation Drawings
28
oxy.IQ User’s Manual
Figure 14: Wiring options (ref. dwgs. 702-285 and 702-286)
Maximum load of analog input device is
dependent on power supply voltage
and cable resistance.
Maximum load of analog input device is
dependent on power supply voltage
and cable resistance.
1. For class 1 div 2 installation, the oxy.IQ shall be installed in
an appropriately rated final enclosure accepting a division 2
wiring method per NEC/CEC.
704-1317-xx
(black jacket)
Class 1 div 2 enclosure
Hazardous location
Non-hazardous location
blue
brown
brown
+24V
24V return
Power supply
4-20MA analog
input device
704-1317-xx
(black jacket)
blue
brown
brown
4-20MA analog
input device
+24V
24V return
Power supply
Non-hazardous location
Hazardous location
Non-hazardous location
IS cable
704-1318-xx
(blue jacket)
+24V
24V return
Power supply
4-20MA analog
input device
R1
R2
ZBB bus bar
MTL7706
BLACK
black
green
blue
brown
IS ground
Non-IS cable
red
green
1
3
2
4
1. Equipment connected to barrier
inputs must not use or generate
more than 250V.
2. Total load of R1 + R2 must equal
250 ohms ±5%.
oxy.IQ transmitter
oxy.IQ Transmitter
oxy.IQ transmitter
Содержание Panametrics oxy.IQ
Страница 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Страница 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Страница 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Страница 38: ...Appendix A Outline and Installation Drawings 30 oxy IQ User s Manual no content intended for this page...
Страница 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Страница 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Страница 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Страница 55: ......