Chapter 2. Installation
6
oxy.IQ User’s Manual
Figure 3: Packaged Oxygen Sensor
2.
Remove the sensor manifold by unscrewing it from the blue knurled nut on the sensor base at the bottom of the
electronics module.
IMPORTANT:
The maximum operating sample pressure for the oxy.IQ is 10 psig except when the ambient air adapter
is used for the sample manifold. The burst pressure of the sample manifold is 200 psig for all bases
except the ISO KF-40 flange adapter. Since this option is designed for a vacuum flange it cannot handle
sample pressure greater than 20 psig. Be sure the sample conditioning system is designed to maintain
the oxy.IQ pressure below these limits, and that the oxy.IQ outlet is vented to atmosphere during
operation and calibration.
3.
Using PTFE tape as a sealant, connect the sample gas inlet and outlet to the 1/8” NPT ports on the sensor manifold
below). Either port may be used as the inlet or the outlet, as the direction of flow does not matter.
Figure 4: Sensor Manifold Installation
Sample Inlet
Sample Outlet
Sensor Manifold
PTFE Tape
Содержание Panametrics oxy.IQ
Страница 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Страница 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Страница 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Страница 38: ...Appendix A Outline and Installation Drawings 30 oxy IQ User s Manual no content intended for this page...
Страница 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Страница 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Страница 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Страница 55: ......