Chapter 3. Initial Setup and Operation
12
oxy.IQ User’s Manual
3.3.3.2
Recalibrating an Existing Sensor
For an existing sensor, continue the air calibration procedure as follows:
1.
Press the
Enter
key to select the
NO
menu option.
2.
As instructed, remove the sensor manifold to expose the oxygen sensor to ambient air for about two minutes.
Then, press the
Enter
key to continue.
3.
A message indicating that the calibration is in progress will be displayed, and then the calibration data will be
shown. At that time, press the
Enter
key to save the calibration data and return to measurement mode.
3.3.4 Span Gas Calibration
Before beginning the span gas calibration, make sure the oxy.IQ is indicating an O
2
level less than the span gas value,
to ensure an accurate calibration. Then, start the flow of the span gas to the sensor. For accurate calibration, the
span gas should have an oxygen content of 70-90% of the range being calibrated.
To perform the span calibration, complete the following steps:
1.
Use the equation below to calculate the expected mA output that corresponds to the known oxygen content of
the span gas:
For example, if the span gas contains 80 ppm oxygen and the 0-100 ppm range is being calibrated, the analog
output should equal 4 + 16 x (80/100) = 16.8 mA.
2.
If you have not done so already, start the flow of span gas to the sensor and allow both the 4-20 mA output
reading and the display reading to stabilize.
3.
After the reading has stabilized, press the
Enter
key to enter the
Main Menu
.
4.
Press the
Enter
key to enter the
Calibration
menu.
5.
Press the
key and then press the
Enter
key to select the
Span Gas
menu option.
6.
Press the
and
keys until the measurement agrees with the span calibration gas value.
7.
Confirm that the reading on the display has stabilized, and press the
Enter
key to save the calibration. Then, press the
Cancel
key twice to return to measurement mode.
4.0 16.0 Span Gas ppm
Full Range ppm
----------------------------
+
mA Output
=
Содержание Panametrics oxy.IQ
Страница 1: ...panametrics com BH024C11 EN G oxy IQ Panametrics Oxygen Transmitter User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 12: ...Chapter 1 Features and Capabilities 4 oxy IQ User s Manual no content intended for this page...
Страница 26: ...Chapter 5 The Service Menu 18 oxy IQ User s Manual no content intended for this page...
Страница 30: ...Chapter 6 Specifications 22 oxy IQ User s Manual no content intended for this page...
Страница 38: ...Appendix A Outline and Installation Drawings 30 oxy IQ User s Manual no content intended for this page...
Страница 42: ...Appendix C Order String 34 oxy IQ User s Manual no content intended for this page...
Страница 50: ...Appendix E Cell Models 42 oxy IQ User s Manual no content intended for this page...
Страница 54: ...Warranty 48 oxy IQ User s Manual no content intended for this page...
Страница 55: ......