The flow rate is again outside the original ±10%
Hysteresis
window, but the data exists for a time period greater than the
Bad Data Rejection w
indow . In this instance, the meter interprets the data as a new valid flow rate and moves the
Hysteresis
window to correspond with the new established flow rate .
New
±10% Hysteresis
Limit
Old
±10% Hysteresis
Limit
Bad Data
Rejection Window
150
100
50
0
110
90
Flow
1
2
3
4
4 Samples Outside
Hysteresis Limit
Flow Outside Original
Hysteresis Limit
Figure 71: New valid flow data
Energy Page
Figure 72: Energy units selection
Power Units
Watts
Btu per hour (Btu/hr)
Thousand Joules per hour (kJ/hr)
Kilowatts (kW)
Thousand Btu per hour (MBtu/hr)
Million Joules per hour (MJ/hr)
Megawatts (MW)
Million Btu per hour (MMBtu/hr)
Table 15: Power unit choices
Energy Units
British Thermal Units (Btu)
Ton
Thousand Watt Hours (kWh)
Thousands of Btus (MBtu)
Thousand Joules (kJ)
Million Watt Hours (MWh)
Millions of Btus (MMBtu)
Million Joules (MJ)
Table 16: Energy unit choices