Close
Store to Pipe Wall Setting
0.422 in
X -
X -
X
X
Y
Y
Y
Y
35000
-35000
-21000
21000
7000
-7000
30
45
60
75
90
105
120
135
150
15
0
Remove from Pipe and Clean Transducer
THEN Check to Lock Reference
Automatic Analysis Mode
Ultrasonic Thickness Gauge
Aluminum OD=1.4 in
01P Thin Metal Pipe
Thickness Waveform
Thickness Envelope
Best Peak
0.387 in
Peak 1
0.319 in
Peak 2
0.500 in
PkM
Rpk
Pk1
Pk2
Figure 58: Pipe thickness gauge wizard screen
3 . Before applying couplant and putting the transducer on the pipe, wait for the waveform to stabilize (about 2 . . .4 seconds) .
If measuring pipe with wall thickness greater than about 0.1 in. or 2.5 mm:
4 . Press the
Remove from Pipe and Clean Transducer
check box . A checkmark appears in the check box and the
Red-Rpk
line in the waveform display stops moving . This locks the reference .
5 . Apply couplant to pipe/transducer and apply the transducer to the pipe so that the cable is perpendicular to the pipe (as
well as the alignment line on the bottom of the transducer) . The waveform and measurements settle in a few seconds .
Pipe Thickness
Transducer
5MHz
08
Bottom View
of
Pipe Thickness Transducer
Pipe
Centerline
Place Across
Pipe Centerline
Figure 59: Thickness gauge transducer positioning
6 . The inside of the green button in the lower right of the screen shows the pipe wall thickness measurement .
7 . Lastly
Press
the green measurement button to store pipe thickness to
Setup > Pipe > Wall Thickness.
If measuring copper tubing or other types of metal tubing with wall thickness less than 0.1 in. or 2.5 mm:
7 . Apply couplant to pipe/transducer and apply transducer to pipe so the cable is perpendicular to the pipe (as is the notch
in the transducer) . The waveform settle in a few seconds .
Setup Group
Page 45
November 2016
HYB-UM-00090-EN-04