LOG (DTTT LOGGING) GROUP
Main
Adv
Cal
System
Setup
Meter
I/O
Log
Select Data
Setup
Figure 73: Log Group
Setup Page
The setup screen controls the selection of the Logging Rate and has a software button for starting and stopping logging
sessions .
Figure 74: Datalogging setup
Data Logger Control
MPORTTNT
I
To enable datalogging, a site from the Setup > Site page must be created or selected from previous site names. If a site is not chosen,
the datalogger start/stop control button does not function and the control is grayed out.
Logging Rate
The
Logging Rate
entry tells the flow meter how often to collect data points . The logger memory area has the capacity to store
more than 300 individual files consisting of a maximum of 65,500 points per log file .
The amount of time the logger collects data depends on the
Logging Rate
programmed into the flow meter . In general, the
logging time is calculated by dividing 65,563 by the number of data points recorded per minute .
Logging Time (Minutes) =
65,563
Number of Samples per Minute
Log (Data Logging) Group
Page 57
November 2016
HYB-UM-00090-EN-04