Operation
Hi Flow
Sampler
10
Instruction 0055-9017
IMPORTANT
:
Avoid sampling leaded gasoline vapors, or gases or vapors that contain
silicones or sulfur compounds. Tetraethyl lead, silicones, and sulfur compounds can form
contaminating compounds on the sensor element (poison the sensor), with resulting loss
in sensitivity.
Always purge the instrument with clean air after testing. This removes combustibles
from the sensor chambers and prolongs the life of the sensors.
2.2 Battery Connection
WARNING:
Explosion hazard! DO NOT connect or disconnect the battery in an unsafe
atmosphere.
The instrument’s power cord is located inside the backpack’s right-side pocket. Insert a
battery pack into this pocket and connect it to the power cord.
Figure
2-1
. Battery Pack Connection
2.3 Sampling Hose and Attachment Connections
Connect the sampling hose to the instrument’s leak-gas inlet by clamping the hose to
the inlet as shown in Figure 2-2. Also connect the background gas sampling hose to the
unit’s BACKGROUND inlet.
Figure
2-2
. Hose and Attachment Connections