Operation
Hi Flow
Sampler
38
Instruction 0055-9017
8.
The measurement process continues until
Stop
is selected.
9.
Measurements that were saved can be downloaded to a personal computer at a
later time for analysis. Refer to Section
2.16 Importing Saved Data into a
2.19 Message Line
Various messages will appear at the bottom of the LCD that describe what functions are
currently being performed or errors that have been detected. The following list
describes the meaning of these messages.
Note that detailed explanations of the error messages can be found in Section 5.2
Troubleshooting Guide
.
TABLE
2-3.
Message Line Explanations
Message
Explanation
Calibration Passed
Calibration procedure was successful.
Calibration Failed
Error – The calibration procedure was
not
successful. Refer to Section 5.2.
Saving Record
A test record is currently being saved to memory.
Memory Full
Error – The instrument tried to save a test record with 1000 records already
stored in memory. Refer to Section 5.2.
No Records Saved
An attempt was made to send records to a computer with
no
test records
stored in memory.
Flow Rate Lowered
The flow rate has just been lowered signaling the start of measurement #2
during an automatic or manual leak test. Refer to Section 2.14.
Erasing All Records
All test records are being erased after selecting “Erase records” per
Section 2.15.3.
Check Background Pump
Error – The background-gas pump’s flow rate has been severely reduced. Refer
to Section 5.2.
Check Leak Pump
Error–The leak-gas pump’s flow rate has been severely reduced. Refer to
Section 5.2.