Operation
Hi Flow
Sampler
30
Instruction 0055-9017
1.
Install serial data cable P/N 104-4027 (straight through, 6 foot, DB9 male to DB9
female) between the computer’s COM port and the Hi Flow
COMPUTER
connector. (For computers having a 25-pin or USB COM port, the user must supply
the appropriate adapter, or use a serial cable with appropriate connectors on each
end.)
2.
On the computer, start HyperTerminal by doing one of the following:
- If HyperTerminal has already been set up to communicate with the Hi Flow
Sampler, then select
Start
>
Programs
>
Accessories
>
Communications
>
HyperTerminal
, and double-click the filename or icon that has been associated
with the instrument.
- If HyperTerminal has never been run in association with the Hi Flow
Sampler, then a new
HyperTerminal connection needs to be made as follows:
a. Select
Start
>
Run
. In the Run box type “hypertrm.exe
”
and click
OK.
The
New Connection
dialog box should appear.
b. Type in a New Connection Name such as “HiFlow”. Click
OK
to select the
default icon (if desired, select a different icon before clicking OK). The
Connect To
dialog box should appear.
c. Click the
Connect using
drop-down menu and choose the
COM
port to which
the Hi Flow
is connected. Then click
OK
to display the
COM
Properties
dialog box for that COM port.
d. Set Port Settings to:
- Bits per second:
115200
- Data bits:
8
- Parity:
None
- Stop bits:
1
- Flow control:
None
Then click
OK
to display HyperTerminal’s main window.
e. Click
File > Save
. This creates a file with an .ht extension using the name that
was entered in Step b.
3.
After HyperTerminal is running: click
Transfer
>
Capture Text
>
Start
to accept the
default drive\directory\filename to which all received data will be stored as an ASCII
text file.
Note:
The captured text file will be stored on the hard drive in the folder shown in the
File dialog box. If desired, click
Browse
to select a different drive, directory and filename.
The next time data is captured, it will be stored under the same location and filename.
4.
From the High Flow Sampler’s Expanded Main Screen, start the download process by
selecting
Menu
>
Access
Test IDs
>
Send
Test IDs
.