Operation
Hi Flow
Sampler
12
Instruction 0055-9017
2.5.1 Flange Strap (34", 80", and 137")
Flange leaks can be captured by wrapping this attachment around the edge of the
flange, and securing in place using the two Velcro strips.
Attach the Hi Flow
Sampler’s main sampling hose to the flange strap by simply pushing
the end of the hose into the attachment’s hose coupling. A click should be heard when
the hose has been successfully connected. Release the hose by pushing in the coupling’s
release mechanism.
Figure
2-4.
Flange Strap
2.5.2 Beveled Nozzle Tool (6.5" and 24")
For simple pin-hole type leaks, position this attachment directly over the leak source.
Figure
2-5.
Beveled Nozzle Tool
2.5.3 Capture Bag
This reusable capture bag measures 36" x 36" and can be used to completely surround a
component that may have multiple leak sources, or contains a leak source that is
difficult to pinpoint or reach.
Attach the Hi Flow
Sampler’s main sampling hose to the bag by simply pushing the end
of the hose into the bag’s hose coupling. A click should be heard when the hose has
been successfully connected. Release the hose by pushing in the coupling’s release
mechanism.
Use the draw strings to close up the end of the bag. DO NOT, however, completely close
off the bag – allow air to flow into the bag so as to replace the volume of gas and air
being removed during the test.