50
Type:
Common Command or Query
Syntax:
*ESE<ws><NRf>
Examples: *ESE 48 (Enables the CME and EXE bits)
*ESE 255 (Enables all standard events)
Query
Syntax:
*ESE?
Response: <NR1>
c)
*ESR?
- Standard event status register query
This query is used to read the value of the Standard Event Status Register. Reading the register
clears it.
Type:
Common Query
Syntax:
*ESR?
Response: <NR1>
d)
*PSC
- Power on status clear command
This command is used to control the automatic power-on clearing of certain status functions.
Arguments
Type:
Boolean
Type:
Common Command or Query
Command
Syntax:
*PSC<ws><Boolean>
Examples: *PSC ON or *PSC 1
*PSC OFF or *PSC 0
Query
Syntax:
*PSC?
Response: ASCII 0 for OFF
ASCII 1 for ON
When set to ON (1), the Service Request Enable Register and the Standard Event Status Enable
Register are cleared on power-on.
e)
*SRE
- Service request enable command
This command sets the Service Request Enable Register bits.
Arguments
Type:
NRf
Range: 0 to 255. Non integer arguments are rounded before execution.
The value of bit 6 is ignored, and is set always to zero.
Type:
Common Command or Query
Syntax:
*SRE<ws><NRf>
Examples: *SRE 48 (Enables reporting of ESB and MAV events)
Query
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30