51
Syntax:
*SRE?
Response: <NR1>
f)
*STB?
- Status byte query
This query is used to read the value of the Status Byte.
Type:
Common Query
Syntax:
*STB?
Response: <NR1>
The value of the Status Byte read with the *STB? query may differ from that read with the Serial
Poll. Bit 6 of the STB will be set as long as a reason for requesting service exists, while bit 6 of
the STB as read by the Serial Poll is cleared by the Serial Poll.
Device Trigger Commands
a)
*TRG
- Trigger command
This command is analogous to the IEEE 488.1 Group Execute Trigger interface message, and has
the same effect. It is used to trigger the device to output a wave, and is accepted only when the
trigger mode is set to Trigger, Gate or Burst, and the trigger source is set to BUS.
Type:
Common Command
Syntax:
*TRG
Stored Settings Commands
a)
*RCL
- Recall instrument state
This command is used to restore the state of the device to that stored in the specified memory
location.
Arguments
Type
<NRf>
Range
0 to 49. Non integer values are rounded before execution
Type:
Common Command
Syntax:
*RCL<ws><NRf>
Example: *RCL 0 (Recall default state)
*RCL 49
Stored setting location 49 stores the last instrument setting before power down.
b)
*SAV
- Save instrument state
This command is used to store the current instrument state in the specified memory location.