65
Arguments
Type:
Numeric.
Units:
MHz, KHz, Hz (default)
Range:
Dependent on the frequency range of the current function.
Rounding:
The value is rounded to 4 digits.
Command Type:
Setting or Query
Setting
Syntax:
[:SOURce:]SWEEP:STOP<ws><frequency>[units]
[:SOURce:]SWEEP:STOP<ws>MINimum|MAXimum
Examples:
SWEEP:STOP 5KHZ
SWEEP:STOP 5E3
SWEEP:STOP MAXIMUM
SWEEP:STOP MIN
Query
Syntax:
[:SOURce:]SWEEP:STOP?[<ws>MAXimum|MINimum]
Examples:
SWEEP:STOP ?
SWEEP:STOP ? MAX
Response:
NR3
Phase Adjust
[:SOURce:]PHASe<ws><phase>
This command controls the phase adjustment
Arguments
Type:
Numeric
Units:
one (degrees implied)
Range:
-180 to +180 (other values converted to this range)
Rounding:
to integer
Command Type:
Setting or Query
Setting
Syntax:
[:SOURce:]PHASe<ws><phase>
[:SOURce:]PHASe<ws>MINimum|MAXimum
Examples:
[:SOURce:]PHASe 500
Query
Syntax:
[:SOURce:]PHASe?[<ws>MINimum|MAXimum]
Response:
NR3
Pulse Setting
The following commands control the pulse function:
Note that width + 0.6 x (rise + fall) < period in order to have valid values.
PULSe PERiod
This command sets the pulse period to the specified value.
Note:
This also sets the pulse frequency.
www.calcert.com
1.800.544.2843
0
5
10
15
20
25
30