72
:STARt <numeric value>
:LENGth <numeric value>
:MARKer
[:ADDRess] <numeric value>
:STATe <Boolean>
:LENGth <numeric value>
:SAVe <numeric value>
:LOAD <numeric value>
Point Rate
ARBitrary:PRATe <point rate>
This command is used to set the point rate. It is coupled with the frequency of the waveform by the
relation:
𝐹𝑟𝑒𝑞𝑢𝑒𝑛𝑐𝑦 =
1
(𝑃𝑜𝑖𝑛𝑡 𝑅𝑎𝑡𝑒 × 𝑊𝑎𝑣𝑒𝑙𝑒𝑛𝑔𝑡ℎ)
Thus changing the point rate will result in a change in frequency.
Arguments
Type:
Numeric
Units:
S, mS, uS, nS
Range:
5nS to 100S
Rounding:
to 4 digits
Command Type:
Setting or Query
Setting
Syntax:
:ARBitrary:PRATe<ws><point rate>[units]
:ARBitrary:PRATe<ws>MINimum|MAXimum
Examples:
:ARB:PRAT 100NS
Query
Syntax:
:ARBitrary:PRATe?[<ws>MINimum|MAXimum]
Response:
NR3
Address
:ARBitrary:ADDRess <address>
This command sets the current address of the waveform. It is used to determine to where arbitrary
data are to be written.
Arguments
Type:
Numeric
Range:
1 to 16,777,216
Rounding:
to integer value
www.calcert.com
1.800.544.2843
0
5
10
15
20
25