57
Rate increments
0.1 - 99.99 ml/h in increments of
0.01 ml/h
100.0 - 999.9 ml/h in increments of
0.1 ml/h
1000.0 - 1200 ml/h in increments of
1 ml/h
Accuracy of bolus infusion
typ. ± 5 % as of a bolus volume > 1 ml
KVO-rate
Delivery rate 10 ml/h: KVO-rate 3 ml/h
Delivery rate < 10 ml/h: KVO-rate 1 ml/h
Delivery rate < 1 ml/h: KVO-rate =
set rate (default setting 0.1 ml/h)
Computer connection
USB connection in combination with
B. Braun interface lead CAN SP
(8713230) including electrical
insulation. Please pay attention to
safety notices.
Air detector
Technical sensitivity:
Detection of air bubbles
0.01 ml
Alarm triggering:
At an air bubble size of typ.
0.02 - 0.3 ml* respectively 1.5 ml/h**
(cumulated value over 1 h from air
bubbles size 0.01ml).
Sensitivity upstream sensor
9 levels from -0.12 bar to -0.21 bar
(pressure reduction)
Occlusion alarm pressures
9 levels up to 1.2 bar
Mechanical occlusion pressure limit
under fault conditions
Occlusion alarm pressure max. 2.1 bar
(210 kPa).
Max. bolus volume 2 ml.
History protocol
1000 last history entries.
100 events for system diagnose.
Refer to separate documents of the
History Viewer for closer information.
* to be set via the service program in increments
of 0.01ml
** to be set via the service program from
0.5-3.8 ml/h in increments of 0.1ml
Chapter 8
TECHNICAL DATA
Level 1
Level 5
Level 9
[bar]
Time to occlusion alarm [min] at rate
Max. bolus
Occlusion pressure
[1 ml/h]
typ. 0.3
09:07
typ. 0.7
typ. 1.2
25:53
46:50
[25 ml/h]
00:33
01:14
02:06
[100 ml/h]
00:07
00:15
00:24
[ml]
0.0347
0.0987
0.1787