OPERATION
2.) Infusion with volume limit
Enter rate and VTBI: The infusion time will be calculated and displayed on the
bottom of the display.
Target: VTBI
3.) Infusion with time limit
Enter rate and time: The infusion volume will be calculated and displayed on the
bottom of the display.
Target: Time
Changing already entered values of VTBI and time
(rate, VTBI and time already exist
at the point of change):
a) Target symbol is placed in front of VTBI:
•
Change of VTBI => Adjustment of time. Old and new target: VTBI
•
Change of time => Adjustment of rate. Old and new target: VTBI
b) Target symbol is placed in front of time:
•
Change of time => Adjustment of VTBI. Old and new target: Time
•
Change of VTBI => Adjustment of time. New target: VTBI
1.3 Bolus Application
There are three ways to administer a bolus:
1.) Manual Bolus: Press
n
b
. Then press
k
and hold button. Fluid is
administered for as long as the button is held down.. The infused bolus
volume is displayed.
The max. bolus time is limited to 10 sec.
Reaching this limit is indicated by an acoustic signal.
2.) Bolus with volume preselection: Press
n
b
. Then press
l
and set bolus dose
limit by using
q
. Press
n
b
to confirm and start bolus. Depending on the
service tool settings an acoustic signal will sound after finishing the bolus
volume.
3.) Bolus with rate calculation: Press
n
b
. Then press
l
and set bolus dose by
using
q
. Press
k
to confirm bolus dose. Set time with
q
in which a
bolus is to be delivered. Calculated bolus rate is shown on top of the display.
Press
n
b
to confirm and start bolus.
After pressing the button
n
b
, the bolus unit can be selected by using
d
. The
selected unit will be stored and offered as default later on. By this also in dose
mode it is possible to administer a bolus in ml.
You can use the service program to enter a default and a maximum bolus rate.
Once a new therapy is started the device always returns to the default rate
- even if the bolus rate was manually changed beforehand.
Note:
If the bolus limit is not entered after pressing
n
b
, the pump switches
back into the run display automatically.
Chapter 1