Activation of the function:
•
Open data lock in Options Menu with
l
.
•
Select between level 1, 2 or 3 (if activated) with
l
and
r
and confirm
with
k
.
•
Enter code with
q
and press
k
in order to activate data lock.
Changes to the protected values and the bolus function which are marked with
y
are only possible after entering the code. After 20 sec in the Main Menu,
Status Menu, Special Functions Menu and Options Menu the lock will be acti-
vated again. If the wrong code is entered twice the pump will switch into the
last menu. If the wrong code is entered twice again the pump will go into an
audible alarm, a nurse call will go off and the yellow LED blinks. If a target
value was reached while data lock is active a new start of the pump is only
possible after entering the code.
In order to deactivate the function, select “Off” in the data lock, press
k
, enter
the code and press
k
again.
4.3 Bolus Rate
•
Open bolus rate in Options Menu with
l
.
•
Change bolus rate with
q
and confirm setting with
k
.
Note:
Set bolus rate according to therapy requirements. Take care not to overdose!
Given a bolus rate of 1200 ml/h, e.g. 0.33 ml are reached within just one second.
4.4 KVO-Mode
The pump can continue the infusion with a preset KVO-rate after a preselected
VTBI/time has passed (see "Technical Data"). The duration of the KVO delivery is
selected in the service program.
45
OPTIONS
Chapter 4
Event
Change of disposable
Level 1
Level 2
Level 3
Change of parameters
O
O
O
O
Stop of infusion
%
Switching off pump / Standby
O
O
%
PCA bolus with pump-based bolus button
= possible |
O
= not possible |
%
= followed by standby-alarm
O
O
Customisable screen
O
O
Acoustic feedback of demanded boli
O
O
Indicates denied PCA boli
Start of infusion
O
with code for
level 1/2
O