![B. Braun Infusomat Space Скачать руководство пользователя страница 50](http://html.mh-extra.com/html/b-braun/infusomat-space/infusomat-space_instructions-for-use-manual_461302050.webp)
50
“Battery cover removed“
The battery cover is not properly engaged on the
battery compartment.
When pushing on the battery cover listen for “click”.
“Standby time expired“
The set standby time has ended.
Set new time or continue with previously set therapy.
“No battery inserted“
It is not possible to use the pump without a
battery pack.
Turn off pump and insert battery pack according
to description “Overview Infusomat® Space“.
“Drive blocked“
Stepper motor does not deliver due to excess pressure
in the system.
Interrupt patient connection and reinsert the line.
“Calibrate device“
Pump calibration data have changed (e.g. after an
update).
Recalibrate device via the service program.
“Drop sensor connection“
Contact to the drop sensor is interrupted while the
pump is delivering.
Check whether the drop sensor is correctly placed on
the drop chamber. If necessary, replace the drop
sensor or preselect VTBI/time and proceed with
therapy.
“Check upstream“
The upstream sensor triggers an alarm.
Check if roller clamp is closed or infusion line is
kinked. If a drop sensor is connected the upstream
alarm is deactivated.
“Air bubble "/"Accumulated air“ Air inside the system.
Check the line for small air bubbles and disconnect
from patient to repeat priming, if necessary.
“No drops“
The drop sensor does not detect any drops.
The infusion container is empty, the roller clamp is
closed, the drop sensor is not put on, check the line
for obstructions, condensation on drop chamber
(remove by shaking it).
"Too few drops"
The number of fallen drops is lower than the preset
rate.
A negative pressure in a glass infusion container can
be eliminated by opening the vent flap on the drop
chamber. Check whether the infusion bottle is empty,
the roller clamp is completely opened and whether
there are any kinks in the line.
"Too many drops"
The number of fallen drops is higher than the preset
rate.
ALARMS