17
Chapter 1
OPERATION
Note:
The infused volume during bolus with volume preselection counts up.
In order to purge the line at any time while the pump is stopped press
n
b
.
Answer the following question by pressing
u
in order to start the purge
process. Cancel by pressing
k
or any other key.
Caution:
Take care not to overdose! Given a bolus rate of 1200 ml/h, 1 ml will
be administered in just 3 sec. To cancel bolus infusion at any time press
k
.
At low bolus volumes, under dosages due to the start up characteristic of the
pump and the tolerances in the infusion system cannot be excluded. Disconnect
patient while purging.
1.4 Infusion Line Change and New Therapy Start
Note:
Always interrupt the patient connection before changing a line to avoid dosing
errors. Never let the pump run unattended when changing the line. Check and clean
the safety clamp regularly.
•
Press
s
f
to stop the delivery. The green LED goes out. Close the roller clamp
and interrupt the patient connection.
•
Press
x
and open the pump door with
u
. Press down the green opening lever
completely until it locks in place, remove the line and insert a new line.
Note:
If for an unknown reason the pump door cannot be opened anymore, you need
to take a crank from the inside of the battery compartment cover. Use this crank to
remove the emergency aperture cover of the pump. Place the crank in the aperture and
turn it clockwise until the pump door opens.
•
Close the pump door, confirm the inserted line with
k
and open the roller clamp.
•
If required, prime the pump with
u
. Then press
d
to proceed.
•
Establish the patient connection and check the parameters with
t
.
•
Start the infusion by pressing
s
f
.
Note:
A new therapy can be started at any time during a stopped infusion. If the pump
is in the Main, Status or Options Menu, press
C
(repeatedly) and follow the
instructions as described.
1.5 End of Infusion
•
Press
s
f
to stop the infusion. The green LED goes out. Close the roller clamp
and interrupt the patient connection.
•
Press
x
. Answer the question whether the pump door is to be opened with
u
.
•
Press down the green opening lever completely until it locks in place.
Remove the line and close the pump door.