44
Chapter 4
OPTIONS
O P T I O N S
The options functions may be selected and changed while the pump is infusing or
stopped. To edit a menu item, select “Options” in the Main Menu and press
l
.
Then select desired function with
t
and follow the Instructions for Use as described.
4.1 Occlusion Pressure
The higher the pressure level is set at, the higher the pressure level must rise before
triggering an occlusion pressure alarm.
•
Enter pressure in Options Menu by pressing
l
.
•
Choose between nine pressure levels (1=lowest level; 9=highest level) by pressing
l
or
r
and confirm entry with
k
.
4.2 Data Lock
The data lock function protects the device against unauthorized access. A four
digit code (default setting “9119”), which can be changed via the service
program activates this function in level 1 or level 2. There are three security
levels.
Level 1:
A modification of values as well as a bolus application are not possible but a
change of the disposable can be conducted. It is possible to navigate through all
menus and status data can be checked. Starting, interrupting and switching the
pump off is possible.
Level 2:
This level has the same performance characteristic as described under level 1
and additional will not allow the change of disposable. In order to prevent a
data lock alarm the correct code must be entered within 20 sec after the pump
was stopped. Changing the disposable and switching the pump off is only
possible after the code was entered.
Level 3:
This level will allow starting and stopping the pump as well as switching off.
The code for this level may be different for each drug and is defined in the drug
list. A change of the disposable, however, is possible by using the code defined
for the other levels. An overview about the differences between the levels 1, 2
and 3 is given by the following table.