12
Aesculap
®
HiLAN
HiLAN pneumatic motor GA519/GA520R/GA529
6.8
Inspection, maintenance and checks
►
Allow the product to cool down to room temperature.
►
Inspect the product after each cleaning and disinfecting cycle to be
sure it is: clean, functional, and undamaged.
►
Spray through the HiLine motor/handpiece with Aesculap STERILIT oil
spray GB600 with adapter GB600820 until clear oil emerges from the
product.
►
Check the product for any damage, abnormal running noise, overheat-
ing or excessive vibration.
►
Inspect tools for broken, damaged or blunt edges.
►
Set aside the product if it is damaged.
Note
Lubricating the HiLAN pneumatic motors with the original Aesculap oil
spray can lead to oil residues entering the compressed-air tubes. These oil
residues do not harm the materials from which the compressed-air tubes
are made.
6.9
Packaging
►
Follow the instructions for use for the applied packaging and storage
systems (e.g. instructions for use TA009721 for Aesculap Eccos storage
system).
►
Insert the product in its proper position in the Eccos holder, or put it
on a tray in such a way that the product is protected against damage.
Ensure that all cutting edges are protected.
►
Pack trays appropriately for the sterilization process (e.g. in Aesculap
sterile containers).
►
Ensure that the packaging will prevent a recontamination of the prod-
uct.
6.10
Steam sterilization
Note
Regarding materials and temperatures, the HiLAN pneumatic motors have
been designed for steam sterilization. Consequently, do not sterilize HiLAN
pneumatic motors with hot air or ethylene oxide (EO).
HiLAN pneumatic motors can be sterilized in their storage device.
►
Check to ensure that the sterilizing agent will come into contact with
all external and internal surfaces (e.g. by opening any valves and fau-
cets).
►
Validated sterilization process
– Disassemble the instrument
– Steam sterilization through fractionated vacuum process
– Steam sterilizer according to DIN EN 285 and validated according
to DIN EN ISO 17665
– Sterilization using fractionated vacuum process at 134 °C/holding
time 5 min
►
When sterilizing several instruments at the same time in a steam ster-
ilizer, ensure that the maximum load capacity of the steam sterilizer
specified by the manufacturer is not exceeded.
6.11
Sterilization for the US market
■
Aesculap advises against sterilizing the device by flash sterilization or
chemical sterilization.
■
Sterilization may be accomplished by a standard prevacuum cycle in a
steam autoclave.
To achieve a sterility assurance level of 10
-6
, Aesculap recommends the
following parameters:
*Aesculap has validated the above sterilization cycle and has the data on
file. The validation was accomplished in an Aesculap sterile container
cleared by FDA for the sterilization and storage of these products. Other
sterilization cycles may also be suitable, however individuals or hospitals
not using the recommended method are advised to validate any alterna-
tive method using appropriate laboratory techniques. Use an FDA cleared
accessory to maintain sterility after processing, such as a wrap, pouch, etc.
6.12
Storage
►
Store sterile products in germ-proof packaging, protected from dust, in
a dry, dark, temperature-controlled area.
7.
Maintenance
To ensure reliable operation, the product must be maintained at least once
a year.
For technical service, please contact your national B. Braun/Aesculap
agency, see Technical Service.
Aesculap Orga Tray/Sterile container (perforated bottom)
Minimum cycle parameters*
Sterilization
method
Temp.
Time
Minimum drying time
Prevacuum
270 °F/275 °F
4 min
20 min
Содержание Aesculap HiLAN GA519
Страница 2: ...GA523 2 3 m GA464 3 m GA523 2 3 m ...
Страница 3: ......