2.2 Display Screens
2.2.1 Display Screens for 5 Operation Modes and Standby
2.2.1.1 EAL Mode
This mode is for canal measurement. The motor handpiece does not
run in this mode
.
2.2.1.2 CW Mode
The motor handpiece rotates forward 360º, clockwise direction.
2.2.1.3 CCW Mode
The motor handpiece rotates counterclockwise direction only. This
mode is used to inject calcium hydroxide and other medicant. When this
mode is being used, a double-beep sounds continuously.
2.2.1.4 SGP Mode
Safety Glide Path Mode
F: Forward angle, R: Reverse angle
The rotation angle is adjustable, but the forward angle must be equal
to the reverse angle.
2.2.1.5 ATR Mode
ATR: Adaptive Torque Reverse function.
Normal continuous forward rotation, when the load of the file is
greater than the set torque limit, the file will start to rotate forward &
reverse alternately at the set angle.
2.2.2 Torque Display
This appears when the motor is running. Meter shows the torque load
on the file.
Set speed
Set torque
Real time torque
2.2.3 Canal Measurement Display
This appears when a file is inside the canal and the lip hook is in
contact with the patient mouth. Bars in meter show the location of the
file tip. In the EAL Mode, if the length is less than 1.0, the display will
be enlarged.