![B.A. International BA OPTIMA E+ BAE380R Скачать руководство пользователя страница 13](http://html.mh-extra.com/html/b-a-international/ba-optima-e-bae380r/ba-optima-e-bae380r_instructions-for-use-manual_461418013.webp)
22
23
the file would return to normal rotation mode when the load is below the
preset torque value again.
Cautions:
1. Protective function of automatic reverse is ONLY suitable for CW
mode.
2. This function is not available for CCW mode, ATR mode.
3. When the motor handpiece battery indicator indicates a low battery
capacity, the low battery capacity is insufficient for the motor handpiece
to reach the torque value limit, therefore, the auto-reverse function will
not work properly. Please charge it in time.
4. If the motor handpiece is under load all the time, the machine may
stop automatically as a result of overheat protection. If it happens, turn
off the motor handpiece for a while until the temperature drops.
4.7 Motor operation
Please set operation mode, torque and speed as per the recommended
specifications of file manufacturer.
Motor alone mode
When using in motor alone mode,
the torque bar will show on the
screen.
(more information about torque
bar, please see chapter 3.2 Screen
display )
Motor combined with canal
measurement function mode
When using motor combined with
canal measurement function, the
measuring wire must be connected
with motor handpiece by USB
socket, and white socket connects
with patient’s lip by lip hook, keep
the black socket idle.
The canal length indicator bar
will show on the screen (more
information about canal length
indicator bar, please see chapter 3.2
Screen display)
Set parameters of automatic
functions as needed, such as Apical
Action, Auto Start, etc(more
information about automatic
functions, please see chapter 4.3
Parameter setting).
Connection testing
It is strongly recommended to
check the connection testing
every time before use. Touch
the lip hook with the file in the
contra angle and check that all the
bars on the meter on the screen
light up, and the motor should be
reversed continuously, otherwise,
the measuring wire or contra angle
should be replaced.
Load value is lower
than preset torque value
Clockwise rotation
Load value is higher than
preset torque value
Couterclockwise rotation
Load value is lower than
preset torque value again
Clockwise rotation
Содержание BA OPTIMA E+ BAE380R