![B.A. International BA OPTIMA E+ BAE380R Скачать руководство пользователя страница 4](http://html.mh-extra.com/html/b-a-international/ba-optima-e-bae380r/ba-optima-e-bae380r_instructions-for-use-manual_461418004.webp)
4
5
1.7.10 Do not remove the contra angle before the motor handpiece
stops rotating. Otherwise the contra angle and the gear inside motor
handpiece will be broken.
1.7.11 Please confirm that the file is properly installed and locked in
place before starting the motor handpiece.
1.7.12 Please set torque and speed as per the recommended
specifications of file manufacturer.
1.7.13 Error in replacing lithium batteries can lead to unacceptable
risks, so use the original lithium battery and replace the lithium battery
according to the correct steps in the instructions.
1.7.14 Please remove the battery if the motor handpiece is not likely
to be used for some time.
1.7.15 Wireless charging will generate heat, and the surface
temperature of charging base and motor handpiece will rise. It is
recommended not to come in contact with the motor handpiece and
charging base during wireless charging for more than 10 seconds.
1.8 Device safety classification
1.8.1 Type of operation mode: Continuous operating device
1.8.2 Type of protection against electric shock: Class II equipment
with internal power supply
1.8.3 Degree of protection against electric shock: B type applied part
1.8.4 Degree of protection against harmful ingress of water: Ordinary
equipment (IPX0)
1.8.5 Degree of safety application in the presence of a flammable
anesthetic mixture with air, oxygen, or nitrous oxide: Equipment cannot
be used in the presence of a flammable anesthetic mixture with air,
oxygen, or nitrous oxide.
1.8.6 Applied part: contra angle, lip hook, file clip, touch probe.
1.8.7 The contact duration of applied part: 1 to 10 minutes.
1.8.8 The temperature of the surface of applied part may reach
46.6℃.
1.9 Primary technical specifications
1.9.1 Battery
Lithium battery in motor handpiece: 3.7V /2000mAh
1.9.2 Power adapter (Model: UE08WCP-050100SPA)
Input: 100V-240V ~50-60Hz, 400mA
Output: DC5V/1A
1.9.3 Torque range: 0.4Ncm-5.0Ncm
(
4mNm ~ 50mNm
)
1.9.4 Speed range: 100rpm~2500rpm
1.9.5 Wireless charging
Frequency range: 112-205KHz
Maximum RF output power of the product: 11.87dBuA/m@3m
1.10 Environment parameters
1.10.1 Environment temperature: +5℃ ~ +40℃
1.10.2 Relative humidity: 30% ~ 75%
1.10.3 Atmospheric pressure: 80kPa ~ 106kPa
2 Installation
2.1 Basic accessories of product
Motor handpiece (BA182610) Contra angle (BA182611)
Charging base (BA182612) Nozzle (BA182613)
Protective silicon cover (BA182614)
Power adapter (EU: BA182615; UK: BA182616; US: BA182624)
Содержание BA OPTIMA E+ BAE380R